ALFOSC setup for polarimetry
Calcite
Mounting calcite plates in the aperture wheel
For imaging polarimetry only one calcite plate is
needed, the one called "Cal-90", which is per default always mounted in
the aperture wheel.
For spectropolarimetry you need to mount another
calicte plate, found in the Calcite plates drawer in the filter
carousel, and this one has a polarimetry slitlet mounted on its top. Per
default the Pol_1.8" wide slitlet is used (actual width is 1.5"),
unless the observer wants a narrower slit.
Status: From April 2003 to November 2008 the two
calcite plates were mounted at 90 degrees orientation in their holders.
This way the slit was horizontal and spectro-polarimetry done with grisms
in the vertical (default) orientation. In November 2008 this was changed
to a 0 degree orientation of the calcites, needing the slit to be vertical
and the grisms horizontal! The Pol_1.8" slitlet was checked for
alignment with respect to the calcite on the 9th Feb 2010 (AAD). Mounted in
position 0 in the aperture wheel its best alignment with respect to the CCD
was at encoder position 317400. (AAD)
Staff Instructions:
- Handle the calcite plates with care. They are very soft material and
should never be cleaned with any liquids nor rubbed in any way. Remove dust
by a light blow of air only! Don't touch with fingers. When they are not
mounted, the calcite plates are stored in their holders wrapped in optical
paper inside a plastic envelope in the Calcite plates drawer in the
filter carousel.
- If you have to take the calcite plates out of their holders, then
check their orientation as follows. Use the piece of paper in the
calcite drawer of the filter carousel called: Calcite alignment tool,
which contains lines rotated by angles 0, 45 and 90 degrees. Place the
calcite plate (in its holder) such that the largest edge of the holder is
parallel to the lower line of the frame of the plot. If it is mounted
correctly, all lines will appear as double except for the horizontal one
(because it is displaced horizontally). If this is not the case, then
loosen the inner tube by unscrewing the screw holding the tubes together
(need a small hex key), and correct the alignment.
NB! It is crucial that the two calcite plates are both oriented in exactly
the same way (i.e. not even 180 degrees offset), otherwise the slit
aquisition for spectropolarimetry will not work.
- The polarimetry slitlet should be mounted on top of the calcite plate
(on the side where light enters from the telescope), and is fixed to the
calcite plate holder by tape. Align it by looking through the calcite at
the slit, which now will appear as two slits horizontally offset to each
others. Look directly at the calcite such that you see your eye reflected
in the center. The slit is aligned with respect to the calcite plate when
the two slit images are on a line. See below for aligning the slit with
respect to the CCD.
- Mount the calcite with the slit in the aperture wheel. You have to
unscrew all four screws and take out the the whole slitholder. (Take care
that screws don't fall into the instrument!)
- Update the wheel content using the Setup Editor. Now align the slit with
respect to the CCD:
- Put the polarimetry slitlet in the beam and set all other positions
to open.
- Set the internal ALFOSC focus to 1070 (default is 1940).
- Turn on the ThAr lamp in the ALFOSC calib unit.
- Make a small window by typing alfosc.polwin in the sequencer.
- Take an image using for instance exp 1.
- Use imexam to find the x-position of the top and bottom part
of the two slitlets split by the calcite. Measure the x-positions with
the key 'j'.
- If the slit is tilted, change the encoder position of the wheel in
the program EditAlfosc and iterate until the slit is aligned.
About -1400 units is needed to rotate the slitlet clockwise by an amount
corresponding to dx = 3 pixels from upper slit top to lower slit bottom.
Note that the slit will be displaced horizontally on the image when you
change encoder values. This is normal.
- When the slit is vertical, type resetxy and take a full
image to determine the slit position (i.e. the xy values for the center
of the upper slit) in order to update the file
~obs/alfosc/slitpos.def. Instruct the observer to use the upper
image of the target for slit acquisition.
- Make an appropriate window for spectroscopy for the user by setting
xbeg, ybeg, xsize, ysize in a script such as for instance:
#!/bin/tcsh
alfosc.resetxy
alfosc.ybeg 800
alfosc.ysize 300
alfosc.xbeg 700
alfosc.xsize 1348
Retarder plate
Mount 1/2 wave or 1/4 wave plate in FAPOL
The idea is that FAPOL will always be mounted in FASU and the default
retarder plate is the 1/2 wave plate, which is used to measure linear
polarimetry. If you have to change the plate, make sure you handle it
with care. These plates are expensive. They are located in the their
marked boxes in the cupboard above the filter table in the electronics
room.
- Make sure the FAPOL carriage is out of the beam.
- Power off the FASU controller.
- Remove the round connector from FAPOL.
- Disconnect the DB25 connecter on the FASU controller side.
- Unscrew the 4 screws that mounts FAPOL in FASU. The unit is heavy
and will fall down a few mm if not held.
- Slide the FAPOL unit carefully out of its hole and bring it down
to the control room.
- Sit at a table and carefully unscrew the 4 screws with rubber
washers that hold the retarder plate in place. The 1/2 wave plate
can in principle go in at any orientation (standard stars will
measure the zero point of the angle of linear polarisation), but
please align such that the red pen mark aligns with the white zero
on the holder. The arrow points to which side is up. Both the 1/2 and
the 1/4 wave plate should be mounted such that the circular mark on
the plate holder points to the circular mark on the carriage (which
is at the outer end of the carriage).
- For the 1/4 wave plate fix one screw per side of the plate. For
the 1/2 wave plate two screws on two opposite sides is the best way.
- Bring the unit back, slide it in and fix it, connect the cables
again and power on the FASU controller. Don't forget to open the
FASU shutter.
Physically mounting the FAPOL unit un FASU
- There is a free way of 13 mm from the top side of FASU B wheel to the
FAPOL unit. FASU filter holders stick out about 2mm above the wheel. To be
conservative, don't put in thick filters in FASU B which will protrude more
than 9mm out of the filter holders.
- Power off the FASU controller.
- Remove the black plate that is facing towards the staircase.
- Mount FAPOL in the hole and fasten it with the screws.
- Remove the blind round connector to the right of the hole.
- Connect the round connector from FAPOL to the connector to the
right of the hole.
- Connect the DB25 connector to the FASU controller.
- Power on the FASU controller (Don't forget to open the shutter).
Connecting FAPOL on a normal instrument mount
- Connect the network cable to the network switch mounted on the adapter.
You can use whatever port you like.
- Connect the 220V cable to a 220V outlet.
Changing the FITS hdr keyword for the retarder plate
- shutdownobssys alfosc
- Change the user from obs to alfosc by typing su alfosc
in the terminal (if you don't know the password check it
from the retarder plate box).
- Edit the file /home/software/alfosc/src/FAPOL/fapoctrl-socket-db.wsh
- On line 37 you will see the line
set retarder_mounted "lambda/2"
- Edit the "lambda/2" to reflect the mounted retarder plate.
- startobssys alfosc
Misc
Start of the control programme
The control programme for FAPOL is started from any of the lisa X-terminal
windows by giving the command (logged in on sandra):
- /usr/local/bin/fapo-control
Since February-2007 FAPOL starts up together
with the rest of ALFOSC. If for some reason it doesn't, then check the
processes and kill all old FAPOL processes that might be running. Start
up FAPOL manually with the above command
Note that after having moved the calibration mirror, it is normal that FAPOL
initializes when you command carriage to go in. The initialization should
take about 10s, and the total time to get FAPOL in is about 40s. If it hangs,
then interrupt the FAPOL s/w by CTRL-C and power cycle the FASU controller
(remember to open the shutter afterwards). Restart FAPOL s/w.
16 retader plate positions
- For testing purposes there are 16 retader plate positions available
(the step size is 22.5 degrees). Non-UIF positions are not updated in
the UIF, i.e. the retader-plate might have different position than
the UIF indicates, however the FITS-headers are updated in a correct way.
The use of the "hidden" positions is the normal FAPO-script
way e.g. "!fapo-112.5" etc.
Comments to Anlaug Amanda Djupvik
|