FIES read-out times

Some approximate readout times are indicated in the table below. For more info on the FIES detector: CCD13

Amplifier Binning Read-out time (sec)
A or B (single) 1x1 90
A or B (single) 2x2 30
A and B (dual) 1x1 45
A and B (dual) 2x2 15

Notes:

  • The default readmode uses the B amplifier.

  • The A and B amplifier gains and bias levels are not exactly equal, but proper spectral reduction for AB-mode dual readout is possible.

    For Xbin=2 together with AB-mode dual readout the two central CCD columns are not recorded in the image:
    the central echelle order (around 4450-4550Å) is difficult to reduce, and spectral coverage may have a gap in this order.

  • Default binning is 1x1, however for the low-res fiber (R=25000) we advise to use 2x2 binning.


Back to top Last modified: 26-Aug-2009