Upgrade requests for NOTCam

Comments to Anlaug Amanda Kaas

  • Clean-array command wanted. Should consist of purely reading the array 3-5 times (TBD whether 3 is sufficient) with no-autosave and no-autoshow in order to save overhead time.

  • Saving options wanted. Want to be able to exclude the reset frame from being stored. For the mexp exposure mode it should be possible to select which of the N readouts to be saved, possibly only the final average image, and it should be possible to select which of the N readouts to be included in the average image.

  • For the frame command we also want different saving options. The user should be able to select which of the readouts to save, if not all of them. (All must be used in the calculation of linear regression though.)

  • Imexamin in BIAS calculates the FWHM in pixels. Because there is no communication between BIAS and NOTCam wheel status, the seeing can not be given in arc seconds unless knowing which camera is in use. Work around solution: calculate the seeing in arc seconds for both cameras, and display both WF seeing and HR seeing.

  • Have the ability to individually initialise separate mechanisms. DONE!

  • For the exp and the mexp commands, let BIAS store the reset-subtracted end-of-exposure frame as the first image in the cube, and the reset frame as the second. Now it stores the reset frame as the first, and the simple end-of-exposure frame as the last. The suggested modification will make it possible to display the actual image, and probably imexamin under BIAS will work. DONE!

  • For the frame command, let BIAS store the linear regression result as the first frame in the cube. Same reason as above. Now it is the last. Keep the reset frame stored in the last position. Store the individual non-destructive read-outs as reset-subtracted images. DONE!

  • The mexp t N command makes a summed image of all the N reset subtracted images. Make instead an average, otherwise the display goes crazy with the wrap around effect. DONE!

  • Imexamin under BIAS frequently crashes or gives strange values. Currently we have to use the POSTPROCESSING ON option in BIAS in orer to use imexamin, which slows down image aquisition. Also, we have to store all the focus acquisition frames and other crap test frames. DONE!

  • Removable vacuum jacketed extensions to the LN2 fill and vent tubes to limit spilling of LN2, and especially to prevent spilling into the controller box. NB! This is high priority!

  • When BIAS displays a newly acquired image in SAOimage the cuts values are always wrong (probably because they are obtained by finding the histogram of the reset frame?). FIXED!

  • Can data acquisition be speeded up by not letting BIAS display the reset frame? DONE!

  • Improve the information to the FITS headers. Neither array temperature, LN2 temperature, pressure, shutter status, etc. are correct values. Missing KEYWORD: observer. DONE!

  • Let the macro "volt" be run automatically upon BIAS start up. DONE!

  • Generate the correct filename automatically. DONE!

  • Change "CCD" to "IR array" everywhere in the displayed windows. Also in the window headers.