MOSCA quality test instructions
(~30 minutes total time)
 
Taking images
-  Get beta light source from carousel, don't drop!
 -  Put beta light into FASU B, position 6, align with marks on slot edge
 -  Close mirror covers or turn off all lights from dome and close the door
 -  Edit the setup file for FASU B by typing editsetup  on
marissa.
 -  Start MOSCA software, filename should be the normal name for current day in order to save images to CD
 -  Run qc-script, this will:
 
  - take a sequence of  3 darks (not saved), 2 40 second flats (saved), 3 darks (not saved) and 2 darks (saved),
   
 -  analyses the images, if the results are OK, accept them
   
 - then puts 2x2 binning on and takes 3 darks (not saved), 2 10 second flats (saved), 3 darks (not saved) 
    and 2 darks (saved).
   
 -  analyses the images, if the results are OK, accept them
 -  check this for nice pictures  
 
  
 -  Take out beta light from FASU B and put back to carousel
 - Edit again the setup file as above to remove "Beta Light"
 -  The whole process takes rather long time, about 23 +  7 minutes.
  