Table Of Contents:
Staff setup Procedure - Preparing TurPol
General
Cooling of the instrument takes at least 4 hours and it should therefore be
mounted a reasonable time ahead of this.
Mounting Turpol on the telescope
TurPol must be mounted on the adaptor in such an orientation that the
ocular tube is pointing 180 degrees AWAY from the StanCam dewar.
(Follow the marks.) Otherwise the periscope optics for viewing the
diaphragms on TV do not work properly. The correct mounting bolts are
listed inside the telescope fork. (Use the green ones without washers.)
Make sure that the tubes leading to the refrigerator unit are free to
move with the telescope movement and do not snag on anything.
Remember to balance the telescope correctly. Set the counter weights to
position 0.12 by typing on the TCS count-w-hei 0.12. You can check
how well the telscope is balanced at alt 60 (zenith angle 30)
and with open mirror covers o-m-c, and then look at TCS status
page 3, where the currents to the two altitude motors are displayed.
These currents (Altitude Motor 1 and 2) should have approximately equal
absolute values. Then type zenith to put the telescope back.
The default telescope focus is set with the command foc-pos 24450.
The default guide TV focus is set typing tv-foc 640.
Cooling system
Check that the cooling system is operating, and that the indicated
temperature is between -16 and -20 degrees Celsius. The cooler must be
switched on at least 3 or 4 hours before starting the
observations. This is usually done by NOT staff. The two red switches
on top of the compressor on the observing floor under the telescope
are switched to ``on''. In case of malfunctioning, refilling of freon
might be needed. Contact the cooling expert (Carlos Perez) for instructions,
or if not possible, follow the
TurPol Cooler
Manual .
Instrument setup
-
Wave Plate
Normally TurPol is equipped with the /2 retarder, which
provides measurements of linear polarisation only. In order to measure
circular & linear polarisation simultaneously, or to do circular
polarimetry, the /2
plate must be replaced by the /4 plate.
To do this, you have to remove the 4 small bolts (3mm Allen key) from
the upper part of the back cover of the polarimeter, disconnect the
motor cable and carefully slide out the module in which the wave plate is
mounted. Note that to be able to slide the module out, you have to
also remove two of the big instrument mounting bolts on that
side. Take the module into the control room or electronics room to be
able to work on it at a table.
Unscrew and remove the /2 plate from its holder and put the
/4 plate instead
(or vice-versa, of course). The orientation
of the /4 wave plate
optical axis must be such that in the starting
position, defined by just actuating the microswitch, the optical axis
marked with the golden lines in the cylindrical wave plate mounting
must be at 45 degrees to the sides of the rectangular black aluminium
plate supporting the unit. This is most easily done by rotating the
large gear clockwise from this starting position by an amount
corresponding to two full rotations of the small gear (which has a
fiducial mark to be able to do this). In this position, the wave plate
mark should be placed parallel to the sides of the unit, or
``vertical'' when you hold the module with the small gear downwards.
The orientation of the /2 plate should also be parallel to
the sides as above, but with the big gear in the starting position
defined by the microswitch. The surface of the wave plate mount with the
optical axis mark should look upwards, i.e. be visible from the side
of the big gear.
Changing the wave plate should be done by NOT staff only.
-
Selecting Polarimetry or Photometry
Check that the CALCITE plate knob is OUT for
POLARIMETRY and push it IN for sky-chopping PHOTOMETRY.
-
Shutter
Make sure that the SHUTTER is closed, i.e. that the knob is in the OUT
position.
-
Power
Check that the high voltage power supplies have been set to zero
voltage, and if not, set them to zero. Also make sure that they are
switched off. Switch on the CAMAC crate in the control room with the
power switch at bottom left of the unit. Switch on the high voltage
power supplies. Turn the voltage selection knobs to the correct
settings. The UBV supply should have the voltage -1150 V and the RI
supply -1800 V. These positions are marked on the front panels of the
power supplies. Do not to put too high a voltage on the
tubes, as this will cause permanent damage!
CAMAC Crate and tests
CAMAC CRATE CONFIGURATION:
Crate controller K-16
Dataway display
Timer TMP-Module
Polarimeter Interface Module
Counter #2
Counter #1
1 U, 3 V, 2 B, 4 R, (5 I)
N1, N2, NP, NTPM, NDD, NCC
Terminal, To IBM/PC
NCC = MAX. position
NDD - ANY position
NP - ANY position
NTPM = NP + 2
N2 = NP - 2
N1 = N2 - 1
For Observation Mode:
- Install the jumpers inside the NP module in position "1".
- Run PP.EXE from personal directory.
For Self Testing Mode:
- Disconnect the 38-pin ELCO from polarimeter.
- Reinstall both jumpers inside the NP module in position "2".
- Disconnect and short "OUT 5" from NP module.
- Run PP_ DEMO.EXE.
Note! (For all computer men):
PC-CAMAC adapter card use I/O ports 0x340 hex - 0x34F hex and IRQ 9,
don't conflict!
|