Wavefront Measuring using ALFOSC
These are some generalised instructions on doing wavefront sensor (WFS)
measurements using ALFOSC. The usual purpose for doing these measurements
is to determine the aberrations of the telescope primary mirror (M1). To
get reliable data the seeing has to be better than 1.1 arcsecs. The exact
procedure will depend on the specific information required, here is a basic
described on how to acquire WFS data of M1.
Setup
- The WFS-lamp and WFS-mask are found in the filter carousel in a
draw labeled Korhanen-Hartmann.
- Put WFS-lamp (LED) in aperture wheel. Set the stepper motor position
to modulus 40000 (i.e. for position N is (N-1) x 40000)
- Put WFS-mask with singlet lens in grism wheel and set the wheel
position to mod. 40000.
- Put H-Alpha narrow #21 filter (also called `Halp 656_3') in FASU.
- Update the slit.def, grism.def and fasu?.def files
using the GUI editor.
- The following day dismount both the WFS lamp and mask from
ALFOSC.
Calibration
This should be done during the day before the night's observing.
Switch on calibration lamp, little switch on same board as lamp in aperture
wheel (switch position 'I'). Using the B#74 (B_Bes 440_10) filter find the
position of the calibration lamp image on the CCD:
- Window the CCD to 600x600 about the image (e.g. alfosc.xbeg 700,
alfosc.ybeg 700, alfosc.xsize 600, alfosc.ysize 600).
- Do 10 second exposure exp 10. In the Sequencer term use imexamine a
command to get the x, y position of the calibration lamp image.
- Edit the
slitpos.def file at tessa:~obs/alfosc/slitpos.def
and add the coordinates but remember to add the xbeg and ybeg offsets to
the x and y position returned by imexamine. (e.g. x=302, y=333 then enter
1002 1033 in the file).
Put the WFS-mask in and the R#76 (R_Bes 650_13) filter and take say
3 calibration (spot) frames (e.g.mexp 10 3). Ensure the spot image
is centred in the 600x600 window with clear space all around it and
that the spots are not saturated or under exposed.
Switch off the calibration lamp (switch position '0'), you should not need it
again. Remember before you go up to ALFOSC to turn the aperture wheel so
you can reach the WFS-lamp.
The reason for determining the position of the cal lamp image on the
CCD is to ensure the path of the light through ALFOSC is the same for it
and the star. It is necessary to get the position of the image with the
CCD windowed because you get different values with a full frame, and you
operate with a windowed CCD.
Below are typical examples of images you can get. The poor image is due to bad
seeing conditions. Note here the area shown
is only 300x300 pixels where the actual CCD window will be twice this size.
|
|
|
Calibration Image
|
Good Stellar Image
|
Poor Stellar Image
|
Details of Measurement Procedure
The seeing must be better than 1.1 arcsecs to get useful data.
Load catalogue /catalogue/wfs4raf.cat. This catalogue has all the
stars from Mv = 4 to 5 for latitude 29 +/-4 degrees, plus the focus stars
from the Blanks catalogue.
Take measurements at zenith angles 5, 10, 20 ,30 ,40 ,50 and 60 degrees
though it may not be possible at the low declinations due to the high airmass.
The rotator ideally should move as little as possible so use a star as
close to 28 45' (telescope declination) as possible.
We wish the rotator to always be about rot-pos 0, this can be achieved
using the `Following the
parallactic angle' TCS feature. Define the start angle use tcs.Instrument-Parallactic-Angle 0
or just i-d-p 0. The rotator should be in automatic rot-auto,
then use CATALOGUE function 6 Preset w par ang or the Sequencer command
preset-with-parallactic-angle-to-selected instead of the usual
4 Preset to this.
To correctly position the star on the CCD put the H-alpha narrow filter
(in FASU) in series with R filter, take a short exposure windowed image (e.g.
using run wfssmall (xbeg=ybeg=700, xsize=ysize=600, exp [short])).
Use the Sequencer script alfosc.slitoff [slit=<slitno>] to move star on cal.
lamp position, (position of WFS-lamp in aperture
wheel). The alfosc.slitoff [slit=<slitno>] command puts the object under the cursor
at the position entered in the slitpos.def file. For the WFS this
ensures the star light through ALFOSC goes through the same path as the
calibration light.
Remove HaN and put in mask with the same window as calibration lamp.
Make a test exposure (e.g.exp 30) to determine the correct integration
time, then check autosave+ and start a mexp tt nn. Five
integration at each position should be satifactory. An exposure time between
30s and 50s is usually sufficent for magnitude Mv = 5 to 4 stars respectively.
General Observation Procedure
A general step-by-step guide to doing the observations:
1.
|
TCS
|
Load the /catalog/wfs4raf.cat catalogue and using
a focus star check the seeing. If better than 1.1 arcsec continue.
|
2.
|
ALFOSC
|
Always with the ALFOSC R filter selected, aperture
and grism wheel open.
|
3.
|
TCS
|
Set the rotator angle using i-d-p 0
|
4.
|
TCS
|
Preset to a suitable star from the catalogue
using function 7 Preset w par ang.
|
5.
|
TCS
|
Start auto-guiding, a-g.
|
6.
|
ALFOSC
|
Put in HaN #21 (Halp 656_3) filter (in series with the ALFOSC
R filter).
|
7.
|
ALFOSC
|
run wfssmall then imexamine t, to
position the star.
|
8.
|
ALFOSC
|
Remove HaN #21 (Halp 656_3) filter and put in WFS-Mask.
|
9.
|
ALFOSC
|
Make the measurements. autosave+, mexp tt
nn, e.g. mexp 40 5.
|
|
|
Repeat items 4 - 9 for all specified zenith angles
(5, 10, 20, 30, 40, 50 ,60).
|
10.
|
ALFOSC
|
The following day dismount the WFS lamp and replace
the WFS mask with a grism (to balance the wheel), but don't both aligning it,
and update the *.def files.
|
|